Is atrial sensing of ventricular far-field signals important in single-lead VDD pacing?
1998
UNLABELLED In single-lead VDD pacing the atrial sensitivity frequently is programmed to sensitive values. Atrial sensing of ventricular far-field signals should be reduced by differential atrial sensing. The aim of the study was to evaluate the effectiveness of this approach. METHODS The study included 10 patients with a single-lead VDD pacemaker (Thera 8948, Lead 5032). The atrial sensitivity was set to its most sensitive value of 0.18 mV and the telemetered intraatrial EGM was continuously recorded. After atrial tracked ventricular pacing, VVI pacing was performed with pacing rates from 100 to 160 beats/min in steps of 10 beats/min and up to 165 beats/min. The peak-to-peak amplitudes of P…
Implantation Strategy of the Atrial Dipole Impacts Atrial Sensing Performance of Single Lead VDD Pacemakers
2002
WIEGAND, U.K.H., et al.: Implantation Strategy of the Atrial Dipole Impacts Atrial Sensing Perfor-mance of Single Lead VDD Pacemakers. Intermittent atrial undersensing is observed in a considerable percentage of patients with single lead VDD pacemakers. Analyzing the 2-year data of the Saphir Multicenter Follow-Up Study, the authors investigated predictors for the occurrence of undersensing. The study included 194 patients with high degree AV block who received a VDD pacemaker system with an identical sensing amplifier. Placement strategy of the atrial dipole was left to the discretion of the implanting physician. At the final position, atrial potential amplitudes were measured during deep …